TY - JOUR
T1 - Comparison of newer IOL power calculation methods for eyes with previous radial keratotomy
AU - Ma, Jack X.
AU - Tang, Maolong
AU - Wang, Li
AU - Weikert, Mitchell P.
AU - Huang, David
AU - Koch, Douglas D.
N1 - Funding Information:
Supported by the AED OCT National Institutes of Health Grant R01EY018184 and in part by an unrestricted grant from Research to Prevent Blindness, New York, New York, United States.
Publisher Copyright:
© 2016, Association for Research in Vision and Ophthalmology Inc. All rights reserved.
PY - 2016/7
Y1 - 2016/7
N2 - PURPOSE. To evaluate the accuracy of the optical coherence tomography-based (OCT formula) and Barrett True K (True K) intraocular lens (IOL) calculation formulas in eyes with previous radial keratotomy (RK). METHODS. In 95 eyes of 65 patients, using the actual refraction following cataract surgery as target refraction, the predicted IOL power for each method was calculated. The IOL prediction error (PE) was obtained by subtracting the predicted IOL power from the implanted IOL power. The arithmetic IOL PE and median refractive PE were calculated and compared. RESULTS. All formulas except the True K produced hyperopic IOL PEs at 1 month, which decreased at ≥4 months (all P < 0.05). For the double-K Holladay 1, OCT formula, True K, and average of these three formulas (Average), the median absolute refractive PEs were, respectively, 0.78 diopters (D), 0.74 D, 0.60 D, and 0.59 D at 1 month; 0.69 D, 0.77 D, 0.77 D, and 0.61 D at 2 to 3 months; and 0.34 D, 0.65 D, 0.69 D, and 0.46 D at ≥4 months. The Average produced significantly smaller refractive PE than did the double-K Holladay 1 at 1 month (P < 0.05). There were no significant differences in refractive PEs among formulas at 4 months. CONCLUSIONS. The OCT formula and True K were comparable to the double-K Holladay 1 method on the ASCRS (American Society of Cataract and Refractive Surgery) calculator. The Average IOL power on the ASCRS calculator may be considered when selecting the IOL power. Further improvements in the accuracy of IOL power calculation in RK eyes are desirable.
AB - PURPOSE. To evaluate the accuracy of the optical coherence tomography-based (OCT formula) and Barrett True K (True K) intraocular lens (IOL) calculation formulas in eyes with previous radial keratotomy (RK). METHODS. In 95 eyes of 65 patients, using the actual refraction following cataract surgery as target refraction, the predicted IOL power for each method was calculated. The IOL prediction error (PE) was obtained by subtracting the predicted IOL power from the implanted IOL power. The arithmetic IOL PE and median refractive PE were calculated and compared. RESULTS. All formulas except the True K produced hyperopic IOL PEs at 1 month, which decreased at ≥4 months (all P < 0.05). For the double-K Holladay 1, OCT formula, True K, and average of these three formulas (Average), the median absolute refractive PEs were, respectively, 0.78 diopters (D), 0.74 D, 0.60 D, and 0.59 D at 1 month; 0.69 D, 0.77 D, 0.77 D, and 0.61 D at 2 to 3 months; and 0.34 D, 0.65 D, 0.69 D, and 0.46 D at ≥4 months. The Average produced significantly smaller refractive PE than did the double-K Holladay 1 at 1 month (P < 0.05). There were no significant differences in refractive PEs among formulas at 4 months. CONCLUSIONS. The OCT formula and True K were comparable to the double-K Holladay 1 method on the ASCRS (American Society of Cataract and Refractive Surgery) calculator. The Average IOL power on the ASCRS calculator may be considered when selecting the IOL power. Further improvements in the accuracy of IOL power calculation in RK eyes are desirable.
KW - Cataract surgery
KW - Intraocular lens calculation formulas
KW - Optical coherence tomography
KW - Radial keratotomy
UR - http://www.scopus.com/inward/record.url?scp=84978726838&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84978726838&partnerID=8YFLogxK
U2 - 10.1167/iovs.15-18948
DO - 10.1167/iovs.15-18948
M3 - Article
C2 - 27409468
AN - SCOPUS:84978726838
SN - 0146-0404
VL - 57
SP - OCT162-OCT168
JO - Investigative Ophthalmology and Visual Science
JF - Investigative Ophthalmology and Visual Science
IS - 9
ER -