Design and evaluation of a portable smart-phone based peripheral neuropathy test platform

Peter G. Jacobs, Peter Hanaway, Joseph Leitschuh, John Condon, Uma Rajhbeharrysingh, Clara Mosquera-Lopez, Carolyn Guidarelli, Kerri Winters-Stone

Research output: Chapter in Book/Report/Conference proceedingConference contribution

4 Scopus citations

Abstract

In this paper, we describe a novel portable test platform that can be used to test peripheral neuropathy either within a clinic or at home. The system, called the PeriVib, is comprised of (1) a small, custom vibration motor designed to apply a vibration stimulus to the toe with constant pressure to test sensation threshold, and (2) a custom smart-phone app that enables a patient to run a series of functional gait and balance tests. Vibration is applied by PeriVib in two separate modes. The first mode, ramp-up, starts at zero amplitude and increases to a maximum level while the patient indicates when they start feeling the pressure by lifting their finger off the touch-screen on the phone. The second mode, ramp-down, starts at a maximal intensity and decreases in intensity; the patient indicates when they stop feeling the vibration. The smart-phone app determines the patient's threshold by recording the vibration amplitude when they indicate the onset or loss of vibratory sensation, depending on the mode. In both modes, the measurement is repeated five times. In addition to controlling the vibration motor during the vibration test, the smart phone app also enables collection of gait and sway metrics through the use of the accelerometer and gyroscope sensors on the smartphone. The entire set of tests requires approximately 5 minutes to complete and can be done by a patient with minimal instructions from a clinician. In a cohort of 28 subjects with a history of chemotherapy-induced peripheral neuropathy, we compared the PeriVib performance with two established threshold sensing systems: (1) a Biothesiometer device and (2) a tuning fork. We found that the sensation threshold estimated by PeriVib correlated well with the Biothesiometer (R2 of 0.68) but less well with the tuning fork (R2 of 0.15). Functional gait and balance metrics did not correlate with peripheral neuropathy severity.

Original languageEnglish (US)
Title of host publication40th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBC 2018
PublisherInstitute of Electrical and Electronics Engineers Inc.
Pages1-4
Number of pages4
ISBN (Electronic)9781538636466
DOIs
StatePublished - Oct 26 2018
Event40th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBC 2018 - Honolulu, United States
Duration: Jul 18 2018Jul 21 2018

Publication series

NameProceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBS
Volume2018-July
ISSN (Print)1557-170X

Other

Other40th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBC 2018
Country/TerritoryUnited States
CityHonolulu
Period7/18/187/21/18

ASJC Scopus subject areas

  • Signal Processing
  • Biomedical Engineering
  • Computer Vision and Pattern Recognition
  • Health Informatics

Fingerprint

Dive into the research topics of 'Design and evaluation of a portable smart-phone based peripheral neuropathy test platform'. Together they form a unique fingerprint.

Cite this