T1 - Design and evaluation of a portable smart-phone based peripheral neuropathy test platform
AU - Jacobs, Peter G.
AU - Hanaway, Peter
AU - Leitschuh, Joseph
AU - Condon, John
AU - Rajhbeharrysingh, Uma
AU - Mosquera-Lopez, Clara
AU - Guidarelli, Carolyn
AU - Winters-Stone, Kerri
N1 - Funding Information:
This research was supported in part by grant Knight Cancer Institute Foundation Grant from Oregon Health & Science University. The content is solely the responsibility of the authors and does not necessarily represent the official views of OHSU.
Publisher Copyright:
© 2018 IEEE.
PY - 2018/10/26
Y1 - 2018/10/26
N2 - In this paper, we describe a novel portable test platform that can be used to test peripheral neuropathy either within a clinic or at home. The system, called the PeriVib, is comprised of (1) a small, custom vibration motor designed to apply a vibration stimulus to the toe with constant pressure to test sensation threshold, and (2) a custom smart-phone app that enables a patient to run a series of functional gait and balance tests. Vibration is applied by PeriVib in two separate modes. The first mode, ramp-up, starts at zero amplitude and increases to a maximum level while the patient indicates when they start feeling the pressure by lifting their finger off the touch-screen on the phone. The second mode, ramp-down, starts at a maximal intensity and decreases in intensity; the patient indicates when they stop feeling the vibration. The smart-phone app determines the patient's threshold by recording the vibration amplitude when they indicate the onset or loss of vibratory sensation, depending on the mode. In both modes, the measurement is repeated five times. In addition to controlling the vibration motor during the vibration test, the smart phone app also enables collection of gait and sway metrics through the use of the accelerometer and gyroscope sensors on the smartphone. The entire set of tests requires approximately 5 minutes to complete and can be done by a patient with minimal instructions from a clinician. In a cohort of 28 subjects with a history of chemotherapy-induced peripheral neuropathy, we compared the PeriVib performance with two established threshold sensing systems: (1) a Biothesiometer device and (2) a tuning fork. We found that the sensation threshold estimated by PeriVib correlated well with the Biothesiometer (R2 of 0.68) but less well with the tuning fork (R2 of 0.15). Functional gait and balance metrics did not correlate with peripheral neuropathy severity.
AB - In this paper, we describe a novel portable test platform that can be used to test peripheral neuropathy either within a clinic or at home. The system, called the PeriVib, is comprised of (1) a small, custom vibration motor designed to apply a vibration stimulus to the toe with constant pressure to test sensation threshold, and (2) a custom smart-phone app that enables a patient to run a series of functional gait and balance tests. Vibration is applied by PeriVib in two separate modes. The first mode, ramp-up, starts at zero amplitude and increases to a maximum level while the patient indicates when they start feeling the pressure by lifting their finger off the touch-screen on the phone. The second mode, ramp-down, starts at a maximal intensity and decreases in intensity; the patient indicates when they stop feeling the vibration. The smart-phone app determines the patient's threshold by recording the vibration amplitude when they indicate the onset or loss of vibratory sensation, depending on the mode. In both modes, the measurement is repeated five times. In addition to controlling the vibration motor during the vibration test, the smart phone app also enables collection of gait and sway metrics through the use of the accelerometer and gyroscope sensors on the smartphone. The entire set of tests requires approximately 5 minutes to complete and can be done by a patient with minimal instructions from a clinician. In a cohort of 28 subjects with a history of chemotherapy-induced peripheral neuropathy, we compared the PeriVib performance with two established threshold sensing systems: (1) a Biothesiometer device and (2) a tuning fork. We found that the sensation threshold estimated by PeriVib correlated well with the Biothesiometer (R2 of 0.68) but less well with the tuning fork (R2 of 0.15). Functional gait and balance metrics did not correlate with peripheral neuropathy severity.
UR - http://www.scopus.com/inward/record.url?scp=85056622652&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85056622652&partnerID=8YFLogxK
U2 - 10.1109/EMBC.2018.8513100
DO - 10.1109/EMBC.2018.8513100
M3 - Conference contribution
C2 - 30440300
AN - SCOPUS:85056622652
T3 - Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBS
SP - 1
EP - 4
BT - 40th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBC 2018
PB - Institute of Electrical and Electronics Engineers Inc.
T2 - 40th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBC 2018
Y2 - 18 July 2018 through 21 July 2018
ER -