MOSFET sensitivity dependence on integrated dose from high-energy photon beams

  • James A. Tanyi
  • , Shane P. Krafft
  • , Tomoe Hagio
  • , Martin Fuss
  • , Bill J. Salter

Research output: Contribution to journalArticlepeer-review

Abstract

The ability of a commercially available dual bias, dual MOSFET dosimetry system to measure therapeutic doses reproducibly throughout its vendor-defined dose-based lifetime has been evaluated by characterizing its sensitivity variation to integrated/cumulative doses from high-energy (6 and 15 MV) photon radiotherapy beams. The variation of sensitivity as a function of total integrated dose was studied for three different dose-per-fraction levels; namely, 50, 200, and 1200 cGy/fraction. In standard sensitivity mode (i.e., measurements involving dose-per-fraction levels ≥100 cGy), the response of the MOSFET system to identical irradiations increased with integrated dose for both energies investigated. Dose measurement reproducibility for the low (i.e., 50 cGy) dose fractions was within 2.1% (if the system was calibrated before each in-phantom measurement) and 3.1% [if the system was calibrated prior to first use, with no intermediate calibration(s)]. Similarly, dose measurement reproducibility was between 2.2% and 6.6% for the conventional (i.e., 200 cGy) dose fractions and between 1.8% and 7.9% for escalated (i.e., 1200 cGy) dose fractions. The results of this study suggest that, due to the progressively increasing sensitivity resulting from the dual-MOSFET design, frequent calibrations are required to achieve measurement accuracy of ≤3% (within one standard deviation).

Original languageEnglish (US)
Pages (from-to)39-47
Number of pages9
JournalMedical Physics
Volume35
Issue number1
DOIs
StatePublished - 2008

Funding

This work was supported in part by a grant from the University of Arizona Foundation. The authors wish to thank Dr. Anatoly B. Rosenfeld, Centre for Medical Radiation Physics, University of Wollongong, Australia and Dr. Dan M. Fleetwood, Vanderbilt University, for their contribution to our understanding of the behavior of the MOSFET system. The authors would also like to thank Dr. Russell J. Hamilton, University of Arizona, for making available additional resources for the completion of this work.

Funders
University of Arizona Foundation

    Keywords

    • Dose response
    • MOSFET
    • Radiation measurement

    ASJC Scopus subject areas

    • Biophysics
    • Radiology Nuclear Medicine and imaging

    Fingerprint

    Dive into the research topics of 'MOSFET sensitivity dependence on integrated dose from high-energy photon beams'. Together they form a unique fingerprint.

    Cite this