Abstract
Finite element modeling has been used as a method to investigate the voltage distribution within the thorax during high energy shocks. However, there have been few quantitative methods developed to assess how well the calculations derived from the models correspond to measured voltages. In this paper, we present a methodology for recording thoracic voltages and the results of comparisons of these voltages to those predicted by finite element models. We constructed detailed 3-D subject-specific thorax models of six pigs based on their individual CT images. The models were correlated with the results of experiments conducted on the animals to measure the voltage distribution in the thorax at 52 locations during synchronized high energy shocks. One transthoracic and two transvenous electrode configurations were used in the study. The measured voltage values were compared to the model predictions resulting in a correlation coefficient of 0.927 ±b 0.036 (average ± standard deviation) and a relative rms error of 22.13 ± 5.99%. The model predictions of voltage gradient within the myocardium were also examined revealing differences in the percent of the myocardium above a threshold value for various electrode configurations and variability between individual animals. This variability reinforces the potential benefit of patient-specific modeling.
| Original language | English (US) |
|---|---|
| Pages (from-to) | 559-571 |
| Number of pages | 13 |
| Journal | IEEE Transactions on Biomedical Engineering |
| Volume | 42 |
| Issue number | 6 |
| DOIs | |
| State | Published - Jun 1995 |
| Externally published | Yes |
Funding
Manuscript received February 16, 1994; revised January 12, 1995. This work was supported by Grant HL.48814 from the National Institutes of Health. D. B. Jorgenson was with the Center for Bioengineering, University of Washington, Seattle, WA 98195 USA. She is now with Heartstream, Inc., Seattle, WA 98121 USA. P. H. Schimpf and Y. Kim are with the Department of Electrical Engineering, FT-10, University of Washington, Seattle, WA 98195 USA. I. Shen is with the Department of Surgery, University of Washington, Seattle, WA 98195 USA. G. Johnson and G. H. Bardy are with the Department of Medicine, Division of Cardiology, University of Washington, Seattle, WA 98195 USA. D. R. Haynor is with the Department of Radiology, University of Washington, Seattle, WA 98195 USA. IEEE Log Number 9410936.
| Funders | Funder number |
|---|---|
| Author National Institutes of Health National Institutes of Health National Institutes of Health National Institutes of Health The Bev Hartig Huntington's Disease Foundation National Institutes of Health | |
| National Institute of Health National Heart, Lung, and Blood Institute | R01HL048814 |
ASJC Scopus subject areas
- Biomedical Engineering
Fingerprint