Abstract

Angiographic thrombus is associated with poorer procedural and clinical outcomes. We report our experience with the X-Sizer thrombectomy device (EndiCOR Medical) from March 2001 to December 2002. Indications for use in 44 patients (33 male; mean age, 60) included myocardial infarction (27), unstable angina (6), periprocedural thrombosis (2), acute (< 24 hr) stent thrombosis (1), and subacute (> 24 hr) stent thrombosis (8). Three cases involved vein grafts. Deployment was successful in 42/44. Difficulty traversing the stent occurred in 5/9 cases of in-stent thrombosis. Median TIMI flow increased from 1 to 2 (P = 0.01) postthrombectomy. Median final TIMI flow was 3. Complications included dissection (1), perforation, device jam on stent (1), disruption of a covered stent (1), distal macro-embolization (4), and transient no-reflow (5). The X-Sizer thrombectomy device improves TIMI flow but does not always prevent distal embolization. Care is needed when treating in-stent thrombosis.

Original languageEnglish (US)
Pages (from-to)177-182
Number of pages6
JournalCatheterization and Cardiovascular Interventions
Volume63
Issue number2
DOIs
StatePublished - Oct 2004
Externally publishedYes

Keywords

  • Coronary
  • Stent
  • Vein graft

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging
  • Cardiology and Cardiovascular Medicine

Fingerprint

Dive into the research topics of 'Procedural efficacy and complications of X-Sizer thrombectomy in de novo and stented lesions'. Together they form a unique fingerprint.

Cite this