Abstract
Angiographic thrombus is associated with poorer procedural and clinical outcomes. We report our experience with the X-Sizer thrombectomy device (EndiCOR Medical) from March 2001 to December 2002. Indications for use in 44 patients (33 male; mean age, 60) included myocardial infarction (27), unstable angina (6), periprocedural thrombosis (2), acute (< 24 hr) stent thrombosis (1), and subacute (> 24 hr) stent thrombosis (8). Three cases involved vein grafts. Deployment was successful in 42/44. Difficulty traversing the stent occurred in 5/9 cases of in-stent thrombosis. Median TIMI flow increased from 1 to 2 (P = 0.01) postthrombectomy. Median final TIMI flow was 3. Complications included dissection (1), perforation, device jam on stent (1), disruption of a covered stent (1), distal macro-embolization (4), and transient no-reflow (5). The X-Sizer thrombectomy device improves TIMI flow but does not always prevent distal embolization. Care is needed when treating in-stent thrombosis.
| Original language | English (US) |
|---|---|
| Pages (from-to) | 177-182 |
| Number of pages | 6 |
| Journal | Catheterization and Cardiovascular Interventions |
| Volume | 63 |
| Issue number | 2 |
| DOIs | |
| State | Published - Oct 2004 |
| Externally published | Yes |
Keywords
- Coronary
- Stent
- Vein graft
ASJC Scopus subject areas
- Radiology Nuclear Medicine and imaging
- Cardiology and Cardiovascular Medicine
Fingerprint
Dive into the research topics of 'Procedural efficacy and complications of X-Sizer thrombectomy in de novo and stented lesions'. Together they form a unique fingerprint.Cite this
- APA
- Standard
- Harvard
- Vancouver
- Author
- BIBTEX
- RIS