Abstract
We present three cases of severely elevated plasma free hemoglobin (PFH) in pediatric patients on mechanical circulatory support devices at a tertiary pediatric care center. Due to severe levels of PFH in the setting of critical illness with the inability to pursue immediate mechanical device exchange, membrane filtration therapeutic plasma exchange (TPE) was performed, which resulted in a lowering of PFH levels. However, long-term outcomes were heterogeneous across the cases. This case series reviews patient presentation, organ function before and after TPE, and the overall role of TPE as an effective treatment option to decrease severely elevated PFH levels. In doing so, we hope to add to what is known about the use of TPE for mechanical red cell hemolysis and provide guidance on its use in critically ill patients.