An Experience With a LeVeen Ascites Shunt

James Holcroft, Herbert Y. Kressel, Robert Prager, Donald Trunkey, Richard P. Jacobs

Research output: Contribution to journalArticlepeer-review

8 Scopus citations

Abstract

A patient with intractable ascites was treated by placement of the LeVeen peritoneovenous shunt. Shunt failure was caused by malpositioning of the venous limb in the inferior vena cava. This was demonstrated by injection of water-soluble contrast material into the venous limb. Revision of the shunt led to diuresis, weight loss, and reduction of ascites. The physiologic process of the LeVeen shunt is reviewed and we suggest placement of a radiopaque marker into the tubing.

Original languageEnglish (US)
Pages (from-to)302-303
Number of pages2
JournalArchives of Surgery
Volume111
Issue number3
DOIs
StatePublished - Mar 1976

ASJC Scopus subject areas

  • Surgery

Fingerprint

Dive into the research topics of 'An Experience With a LeVeen Ascites Shunt'. Together they form a unique fingerprint.