Evaluation of radixact motion synchrony for 3D respiratory motion: Modeling accuracy and dosimetric fidelity

William S. Ferris, Michael W. Kissick, John E. Bayouth, Wesley S. Culberson, Jennifer B. Smilowitz

Research output: Contribution to journalArticlepeer-review

31 Scopus citations

Abstract

The Radixact® linear accelerator contains the motion Synchrony system, which tracks and compensates for intrafraction patient motion. For respiratory motion, the system models the motion of the target and synchronizes the delivery of radiation with this motion using the jaws and multi-leaf collimators (MLCs). It was the purpose of this work to determine the ability of the Synchrony system to track and compensate for different phantom motions using a delivery quality assurance (DQA) workflow. Thirteen helical plans were created on static datasets from liver, lung, and pancreas subjects. Dose distributions were measured using a Delta4® Phantom+ mounted on a Hexamotion® stage for the following three case scenarios for each plan: (a) no phantom motion and no Synchrony (M0S0), (b) phantom motion and no Synchrony (M1S0), and (c) phantom motion with Synchrony (M1S1). The LEDs were placed on the Phantom+ for the 13 patient cases and were placed on a separate one-dimensional surrogate stage for additional studies to investigate the effect of separate target and surrogate motion. The root-mean-square (RMS) error between the Synchrony-modeled positions and the programmed phantom positions was <1.5 mm for all Synchrony deliveries with the LEDs on the Phantom+. The tracking errors increased slightly when the LEDs were placed on the surrogate stage but were similar to tracking errors observed for other motion tracking systems such as CyberKnife Synchrony. One-dimensional profiles indicate the effects of motion interplay and dose blurring present in several of the M1S0 plans that are not present in the M1S1 plans. All 13 of the M1S1 measured doses had gamma pass rates (3%/2 mm/10%T) compared to the planned dose > 90%. Only two of the M1S0 measured doses had gamma pass rates > 90%. Motion Synchrony offers a potential alternative to the current, ITV-based motion management strategy for helical tomotherapy deliveries.

Original languageEnglish (US)
Pages (from-to)96-106
Number of pages11
JournalJournal of Applied Clinical Medical Physics
Volume21
Issue number9
DOIs
StatePublished - Sep 1 2020
Externally publishedYes

Keywords

  • intrafraction motion
  • radixact
  • synchrony
  • tomotherapy
  • tracking

ASJC Scopus subject areas

  • Radiation
  • Instrumentation
  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'Evaluation of radixact motion synchrony for 3D respiratory motion: Modeling accuracy and dosimetric fidelity'. Together they form a unique fingerprint.

Cite this