T1 - Iterative image reconstruction improves the accuracy of automated plaque burden assessment in coronary CT angiography
T2 - A comparison with intravascular ultrasound
AU - Puchner, Stefan B.
AU - Ferencik, Maros
AU - Maehara, Akiko
AU - Stolzmann, Paul
AU - Ma, Shixin
AU - Do, Synho
AU - Kauczor, Hans Ulrich
AU - Mintz, Gary S.
AU - Hoffmann, Udo
AU - Schlett, Christopher L.
N1 - Publisher Copyright:
© American Roentgen Ray Society.
PY - 2017/4
Y1 - 2017/4
N2 - OBJECTIVE. The purpose of this study was to determine whether use of iterative image reconstruction algorithms improves the accuracy of coronary CT angiography (CCTA) compared with intravascular ultrasound (IVUS) in semiautomated plaque burden assessment. MATERIALS AND METHODS. CCTA and I VUS images of seven coronary a rteries were acquired ex vivo. CT images were reconstructed with filtered back projection (FBP) and adaptive statistical (ASIR) and model-based (MBIR) iterative reconstruction algorithms. Cross-sectional images of the arteries were coregistered between CCTA and IVUS in 1-mm increments. In CCTA, fully automated (without manual corrections) and semiautomated (allowing manual corrections of vessel wall boundaries) plaque burden assessments were performed for each of the reconstruction algorithms with commercially available software. In IVUS, plaque burden was measured manually. Agreement between CCTA and IVUS was determined with Pearson correlation. RESULTS. A total of 173 corresponding cross sections were included. The mean plaque burden measured with IVUS was 63.39% ± 10.63%. With CCTA and the fully automated technique, it was 54.90% ± 11.70% with FBP, 53.34% ± 13.11% with ASIR, and 55.35% ± 12.22% with MBIR. With CCTA and the semiautomated technique mean plaque burden was 54.90% ± 11.76%, 53.40% ± 12.85%, 57.09% ± 11.05%. Manual correction of the semiautomated assessments was performed in 39% of all cross sections and improved plaque burden correlation with the IVUS assessment independently of reconstruction algorithm (p < 0.0001). Furthermore, MBIR was superior to FBP and ASIR independently of assessment method (semiautomated, r = 0.59 for FBP, r = 0.52 for ASIR, r = 0.78 for MBIR, all p < 0.001; fully automated, r = 0.40 for FBP, r = 0.37 for ASIR, r = 0.53 for MBIR, all p < 0.001). CONCLUSION. For the quantification of plaque burden with CCTA, MBIR led to better correlation with IVUS than did traditional reconstruction algorithms such as FBP, independently of the use of a fully automated or semiautomated assessment approach. The highest accuracy for quantifying plaque burden with CCTA can be achieved by using MBIR data with semiautomated assessment.
AB - OBJECTIVE. The purpose of this study was to determine whether use of iterative image reconstruction algorithms improves the accuracy of coronary CT angiography (CCTA) compared with intravascular ultrasound (IVUS) in semiautomated plaque burden assessment. MATERIALS AND METHODS. CCTA and I VUS images of seven coronary a rteries were acquired ex vivo. CT images were reconstructed with filtered back projection (FBP) and adaptive statistical (ASIR) and model-based (MBIR) iterative reconstruction algorithms. Cross-sectional images of the arteries were coregistered between CCTA and IVUS in 1-mm increments. In CCTA, fully automated (without manual corrections) and semiautomated (allowing manual corrections of vessel wall boundaries) plaque burden assessments were performed for each of the reconstruction algorithms with commercially available software. In IVUS, plaque burden was measured manually. Agreement between CCTA and IVUS was determined with Pearson correlation. RESULTS. A total of 173 corresponding cross sections were included. The mean plaque burden measured with IVUS was 63.39% ± 10.63%. With CCTA and the fully automated technique, it was 54.90% ± 11.70% with FBP, 53.34% ± 13.11% with ASIR, and 55.35% ± 12.22% with MBIR. With CCTA and the semiautomated technique mean plaque burden was 54.90% ± 11.76%, 53.40% ± 12.85%, 57.09% ± 11.05%. Manual correction of the semiautomated assessments was performed in 39% of all cross sections and improved plaque burden correlation with the IVUS assessment independently of reconstruction algorithm (p < 0.0001). Furthermore, MBIR was superior to FBP and ASIR independently of assessment method (semiautomated, r = 0.59 for FBP, r = 0.52 for ASIR, r = 0.78 for MBIR, all p < 0.001; fully automated, r = 0.40 for FBP, r = 0.37 for ASIR, r = 0.53 for MBIR, all p < 0.001). CONCLUSION. For the quantification of plaque burden with CCTA, MBIR led to better correlation with IVUS than did traditional reconstruction algorithms such as FBP, independently of the use of a fully automated or semiautomated assessment approach. The highest accuracy for quantifying plaque burden with CCTA can be achieved by using MBIR data with semiautomated assessment.
KW - Accuracy
KW - Coronary CT angiography
KW - Coronary plaque burden
KW - Intravascular ultrasound
KW - Iterative image reconstruction algorithms
UR - http://www.scopus.com/inward/record.url?scp=85016479689&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85016479689&partnerID=8YFLogxK
U2 - 10.2214/AJR.16.17187
DO - 10.2214/AJR.16.17187
M3 - Article
C2 - 28177655
AN - SCOPUS:85016479689
SN - 0361-803X
VL - 208
SP - 777
EP - 784
JO - American Journal of Roentgenology
JF - American Journal of Roentgenology
IS - 4
ER -