Utility of COVID-19 antigen testing in the emergency department

W. Frank Peacock, Karina M. Soto-Ruiz, Stacey L. House, Chad M. Cannon, Gary Headden, Brian Tiffany, Sergey Motov, Kian Merchant-Borna, Anna Marie Chang, Claire Pearson, Brian W. Patterson, Alan E. Jones, Joseph Miller, Joseph Varon, Aveh Bastani, Carol Clark, Zubaid Rafique, Bory Kea, John Eppensteiner, James M. WilliamsSimon A. Mahler, Brian E. Driver, Phyllis Hendry, Eugenia Quackenbush, David Robinson, Jon W. Schrock, James P. D'Etienne, Christopher J. Hogan, Anwar Osborne, Ralph Riviello, Stephen Young

Research output: Contribution to journalArticlepeer-review

8 Scopus citations

Abstract

Background: The BinaxNOW coronavirus disease 2019 (COVID-19) Ag Card test (Abbott Diagnostics Scarborough, Inc.) is a lateral flow immunochromatographic point-of-care test for the qualitative detection of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) nucleocapsid protein antigen. It provides results from nasal swabs in 15 minutes. Our purpose was to determine its sensitivity and specificity for a COVID-19 diagnosis. Methods: Eligible patients had symptoms of COVID-19 or suspected exposure. After consent, 2 nasal swabs were collected; 1 was tested using the Abbott RealTime SARS-CoV-2 (ie, the gold standard polymerase chain reaction test) and the second run on the BinaxNOW point of care platform by emergency department staff. Results: From July 20 to October 28, 2020, 767 patients were enrolled, of which 735 had evaluable samples. Their mean (SD) age was 46.8 (16.6) years, and 422 (57.4%) were women. A total of 623 (84.8%) patients had COVID-19 symptoms, most commonly shortness of breath (n = 404; 55.0%), cough (n = 314; 42.7%), and fever (n = 253; 34.4%). Although 460 (62.6%) had symptoms ≤7 days, the mean (SD) time since symptom onset was 8.1 (14.0) days. Positive tests occurred in 173 (23.5%) and 141 (19.2%) with the gold standard versus BinaxNOW test, respectively. Those with symptoms >2 weeks had a positive test rate roughly half of those with earlier presentations. In patients with symptoms ≤7 days, the sensitivity, specificity, and negative and positive predictive values for the BinaxNOW test were 84.6%, 98.5%, 94.9%, and 95.2%, respectively. Conclusions: The BinaxNOW point-of-care test has good sensitivity and excellent specificity for the detection of COVID-19. We recommend using the BinasNOW for patients with symptoms up to 2 weeks.

Original languageEnglish (US)
Article numbere12605
JournalJACEP Open
Volume3
Issue number1
DOIs
StatePublished - Feb 2022

Keywords

  • Covid-19
  • antigen testing
  • diagnostic devices
  • emergency department
  • nasal swab
  • point of care

ASJC Scopus subject areas

  • Emergency Medicine

Fingerprint

Dive into the research topics of 'Utility of COVID-19 antigen testing in the emergency department'. Together they form a unique fingerprint.

Cite this